PRODUCT
The digital display automatic turret type Vickers hardness tester can meet the basic Vickers hardness tester test requirements. The upgraded digital display micrometer eyepiece can directly read the measurement length, so that the measured value can be obtained more quickly and accurately. The tower can automatically switch the eyepiece and the indenter to avoid damage to the instrument caused by manual error, and also greatly improve the efficiency. The 7-inch LCD touch screen can provide more measurement data
Main functions and features:
1. The optical system designed by optical engineers not only has clear images, but also can be used as a simple microscope, with adjustable brightness, comfortable vision, and not easy to fatigue after long-term operation;
2. The 7-inch touch screen can display the test method, test force, upper and lower limits of hardness value, measured indentation length, hardness value, test force retention time, measurement times and can enter year, month, date, test results and data processing Wait;
3. The system comes with a calibration function, and the input error calibration instrument can be manually adjusted for each test force;
4. The adjustable range of optical system brightness: 0-100;
5. It can store up to 100 sets of test data, and display the average value and test times data on the main interface;
6. The test force unit can be manually selected for mutual conversion, the conversion unit is: (Kgf, N)
7. The main interface can display 4 conversion scales at the same time, and can switch: national standard, American standard and international standard. A total of 18 measurement scales can be converted: HRA, HRB, HRC, HRD, HRE, HRF, HRG, HRH, HRK ,
HR15N, HR30N, HR45N, HR15T, HR30T, HR45T, HV, HK, HBW;
8. Both objective lenses can directly measure the hardness,
9. With diagonal length memory function, the diagonal length will be automatically memorized after shutting down, and it will be measured directly after turning on, without the need for secondary reset.
10. The measuring scale can be converted into HK Knoop hardness;
11. Test data can be directly exported via USB
12. The upper and lower limit measurement range can be set and displayed on the main measurement interface
13. The operating system can be converted between Chinese and English;
14. Optional built-in printer, test data can also be output to U disk via USB for storage and analysis;
15. The HSVT-50Z model is realistic: the measured length can directly confirm the hardness value;
16. HVST-50Z model can realize: the position switch of observation-test-measurement is completed by automatic turret
17. The cast aluminum shell has a stable structure without deformation at one time, and the pure white car paint has a high grade, strong scratch resistance, and it is still bright as new after many years of use;
18. Our company has the ability to develop, design, produce and process by ourselves, and provide parts replacement and maintenance upgrade services for our machines for life
Main purpose and scope of application:
1. Steel, non-ferrous metals, metal foil, cemented carbide, metal sheet, microstructure, carbonization;
2. Carburizing, nitriding and decarburizing layer, surface hardening layer, electroplating layer, coating, heat treatment;
3. Glass, wafers, ceramic materials;
Technical parameter:
parameter name | Parameter data |
Vickers hardness scale | HV1,HV5,HV10,HV20,HV30,HV50 |
display | 5-digit hardness value, 4-digit diagonal length (D1, D2), Hold time, number of tests, average, standard deviation, return |
Test force(Kgf) | 1.0Kgf(9.8N)、5.0Kgf(49.0N)、10Kgf(98.0N)、 20Kgf(196N)、30Kgf(294N)、50Kgf(490N) |
Loading control | Automatic (load/hold/uninstall) |
Test force retention time(s) | 1~99 |
Optical channel | Dual light channels (eyepieces and CCD camera channels) |
Optical system | Objectives: 10×, 20× Eyepiece Magnification: 10×(can be changed to 15X) Total magnification (μm): 100×, 200×(can be changed to 150X,300X) Measuring range (μm): 400 Resolution (μm): 0.125 |
Hardness measurement range | (5-3000)HV |
XY test stand (optional) | Size (mm): 100×100 Travel range (mm): 25×25 Minimum reading (mm): 0.01 |
Maximum height of specimen (mm) | 165 |
Maximum width of specimen (mm) | 130 (distance from the center line of the indenter to the wall) |
Data output | Reserve built-in mini printer interface, (RS232 serial interface) optional installation |
voltage | AC220V/50HZ |
Overall size (mm) | 585×200×630 |
Weight (kg) | 42 |
Standard configuration
Host computer | 1 |
weight | 3 |
Large and medium "V" test benches | 1 |
Horizontal adjustment screw | 4 |
Power cable | 1 |
10X digital microscope eyepiece | 1 |
Vickers hardness block | 2 |
Spare fuse 2A | 2 |
Product Certificate | 1 |
Instruction Manual | 1 |